Advanced Search

Journal Navigation

Journal Home

Subscriptions

Archive

Contact Us

Table of Contents

Click here to sign up for SAGE Journal Email Alerts today!

Journal of Parenteral and Enteral Nutrition
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Saved Citations
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Request Reprints
Right arrow Add to My Marked Citations
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Right arrow Citing Articles via Scopus
Google Scholar
Right arrow Articles by Schilling, S.
Right arrow Articles by Jacobs, B. R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Schilling, S.
Right arrow Articles by Jacobs, B. R.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?

Original Communications

The Impact of Needleless Connector Device Design on Central Venous Catheter Occlusion in Children: A Prospective, Controlled Trial

Susan Schilling, RN*, Darcy Doellman, RN, BSN*, Nancy Hutchinson, RN{dagger} and Brian R. Jacobs, MD{ddagger}

From the * Division of Home Health Care,{dagger} Infectious Diseases, and{ddagger} Critical Care Medicine, Cincinnati Children's Hospital Medical Center, Cincinnati, Ohio

Correspondence: Brian R. Jacobs, MD, Professor of Clinical Pediatrics, Director, Technology and Patient Safety, Division of Critical Care Medicine, Cincinnati Children's Hospital Medical Center, 3333 Burnet Avenue, Cincinnati, OH 45229. Electronic mail may be sent to Jacobs{at}chmcc.org.

Background: Intraluminal occlusion is common in children with central venous catheters (CVCs). Although multiple factors predispose CVCs to occlusion, reflux of blood is frequently implicated. We hypothesized that use of either a single-valve or positive-pressure-valve needleless connector device would reduce CVC occlusion rates in comparison to a standard device. We further hypothesized that saline would be as effective as heparinized saline flush in preventing occlusion and infection. Methods: CVC lumens were prospectively capped with 1 of 3 needleless connector devices in a 4-group design. Group 1 lumens were capped with a standard device, group 2 with a single-valve device, group 3 with a positive-pressure-valve device flushed with heparinized saline, and group 4 with a positive-pressure-valve device flushed with saline. Data were obtained regarding occlusion and infection rates and user satisfaction. Results: Three hundred sixty children with 599 CVC lumens completed the study. Complete occlusion occurred in 19/150 (12.7%) lumens in group 1 in comparison to 2/150 (1.3%) in group 2, 5/149 (3.4%) in group 3, and 6/150 (4%) in group 4 (p < .05). There was a trend toward a 2-fold greater infection rate in group 4. User satisfaction was higher in groups 2, 3, and 4 than group 1 (p < .05). Conclusions: CVCs capped with a single-valve or positive-pressure-valve needleless connector device have lower complete catheter occlusion rates than those capped with a standard device. Heparinized saline flush affords no advantages over saline in reducing occlusion rate; however, there was a trend toward lower infection rate with the use of heparinized saline.

Central venous catheters have become essential in the management of adults and children with acute and chronic diseases. Though these devices are clearly beneficial, they are not without risks. One of the most frequent complications associated with the use of these devices is occlusion, which occurs in up to 25% of central venous catheters.1,2 Historical prospective data at Cincinnati Children's Hospital Medical Center (CCHMC) revealed a 12% catheter occlusion rate (internal data). Catheter occlusion can lead to restricted catheter use and delay or discontinuation of patient therapy, thus compromising the quality of care. In addition, the resulting inability to obtain blood samples through these malfunctioning devices can result in the need for peripheral venipuncture that is often difficult and traumatic in children. Characteristics that contribute to catheter occlusion include catheter composition, catheter size, various patient factors, infusate composition, local catheter use protocols, and the use of connector devices.35 A variety of needleless connector devices has been marketed, with claims that they can decrease central venous catheter occlusion and infection rates by preventing reflux of blood through the catheter tip. Needleless connector devices are currently available with 4 basic designs: standard (no valve), single valve, dual valve, and positive pressure valve. Theoretically, needleless connector devices with an internal valve prevent retrograde movement of fluid and associated complications, including catheter occlusion and infection. A positive-pressure-valve needleless connector device was recently shown to be effective in significantly reducing total catheter occlusion rate in comparison to a standard connector device.6,7 There is, however, little additional published evidence supporting the use of one valve connector device design over another in the prevention of central venous catheter occlusion, infection, or other complications. We hypothesized that the use of either a single-valve or positive-pressure-valve needleless connector device would reduce central venous catheter occlusion rates in comparison to the use of a standard device in children with indwelling central venous catheters. In addition, we hypothesized that the use of saline-only flush would be as effective as heparinized saline flush in preventing occlusion and infection.


    METHODS
 Top
 METHODS
 RESULTS
 DISCUSSION
 CONCLUSIONS
 
Study Design
The CCHMC institutional review board approved the study and waived the requirement to obtain informed consent. The Central Venous Catheter Resource Team (CVCRT) at CCHMC is responsible for the oversight of all catheter-related issues for the institution. Using a prospective, sequential study design, all inpatients with newly placed nontunneled or tunneled central venous catheters in the Pediatric Intensive Care Unit (PICU), Cardiac Intensive Care Unit (CICU) and those patients referred to the CVCRT for peripherally inserted central catheter (PICC) placement were eligible for participation in the study. Four groups were studied. Group 1 consisted of central venous catheters capped with a standard needleless connector device (Interlink, Becton Dickinson and Company, Franklin Lakes, NJ), and a 10-unit/mL flush of heparinized saline was used. This device did not have any internal valve. Group 2 consisted of central venous catheters capped with a single-valve needleless connector device (Alaris Smartsite, Alaris Medical Systems, San Diego, CA) and flushed with 10 units/mL of heparinized saline. Group 3 consisted of central venous catheters capped with a positive-pressure-valve needleless connector device (BD Posi-Flow, Becton Dickinson) and flushed with 10 units/mL heparinized saline. In group 4, central venous catheters were capped with a positive-pressure-valve needleless connector device (BD Posi-Flow) and flushed with saline (0.9% sodium chloride). Each study phase was continued until at least 150 devices were studied in each group. Central venous catheters were placed according to standard CCHMC procedure. Once the catheter was successfully placed, the assigned needleless connector device was attached to the catheter hub using the standard CCHMC nursing procedure. A 3-minute scrub consisting of povidone-iodine (10% aqueous) followed by 70% isopropyl alcohol scrub was used for cleansing needleless connectors before IV system entry throughout the study. For lumens requiring continuous infusions, the needleless connector was placed after the continuous treatment was discontinued, and intermittent therapy was initiated. The needleless connector device was changed weekly or when indicated per CCHMC protocol for the duration of the study. All therapies administered through each of the needleless connector devices were consistent with manufacturer's recommendations relating to dose, dilution, frequency, and rate. Only those devices in place for 3 or more consecutive days were included in the final data analysis. Before the beginning of each study phase, caregivers in each study location were extensively educated on the proper use of the respective needleless connector device. All caregivers had access to a CVCRT RN coinvestigator 24 hours each day for questions and problems. Patients exited the study for any of the following reasons: therapy was complete, the capped lumen was needed for continuous infusion, the needleless connector device was not consistently in place for 3 days, the catheter was removed, or the patient expired. At the end of each phase of the study, caregivers were asked to complete an anonymous user satisfaction survey (Table I) specific to that device.


View this table:
[in this window]
[in a new window]

 
Table I User satisfaction survey

 

Data Collection
Data collected included patient demographics; catheter type, size, length, and number of lumens; catheter duration; type of infusate; placement complications; and catheter complications that occurred throughout the course of therapy. Catheter complications included partial occlusion events (defined as the inability to aspirate but able to flush), complete occlusion (defined as the inability to aspirate or flush), mechanical complications (catheter malfunction, tears, breaks, needleless connector cracks, or other dysfunction), phlebitis, and catheter-related bloodstream infection (CRBSI). Decisions to obtain blood cultures were made by the primary care team according to the patient's clinical status. Positive blood or catheter-site cultures during the study were recorded. CRBSI was defined using standard Centers for Disease Control and Prevention definitions.8,9 Patient data were recorded on written case report forms.

Statistical Analysis
The primary outcome variable was the difference in the complete occlusion rate between each of the 4 groups. This outcome variable was used for sample size calculations. Assuming an 8% incidence of complete occlusion in group 1, a 4% incidence of complete occlusion in group 3, a 2-tailed {alpha} of .05, and a β of .2, it was determined that completed data from 140 needleless connector devices would be required in each group. Assuming a 7% dropout rate after enrollment, the investigators elected to enroll 150 devices per group. Secondary outcome variables included the difference in incidence of mechanical complications, device malfunction, infection rate, phlebitis, and user satisfaction between the 4 groups. Values are expressed as mean ± SEM. Categorical variables were analyzed using {chi}2 or Fisher's exact testing. Continuous variables were evaluated using the Student's t-test. A p value of ≤ .05 was considered significant.


    RESULTS
 Top
 METHODS
 RESULTS
 DISCUSSION
 CONCLUSIONS
 
Overall, 392 children with 781 central venous catheter lumens were enrolled in the study. Thirty-two patients with 182 lumens met early exit criteria, resulting in 360 children with 599 central venous catheter lumens for final analysis. Catheters represented included nontunneled (76%), PICC (23%), and tunneled (1%). Mean patient age was 44.3 ± 3.5 months (53.6% male, 46.4% female). Mean catheter duration was 8.3 ± 0.3 days. Baseline characteristics in the 4 groups are noted in Table II. There were no significant differences detected between any of the study groups for gender and age. Patients in group 1 had a significantly smaller mean catheter size than patients in group 3. Mean catheter duration was significantly less in groups 2, 3, and 4 in comparison to group 1.


View this table:
[in this window]
[in a new window]

 
Table II Baseline characteristics of patients and catheters within each study group

 

Table III displays the differences between the occlusion and complication rates between each group. The complete occlusion rates were significantly greater in group 1 (19/150, 12.7%) in comparison to group 2 (2/149, 1.3%), group 3 (5/150, 3.3%), and group 4 (6/150, 4%) (p < .05). In examining the occlusion rates per 1000 device days, there was a 2.1- to 7.4-fold reduction in complete occlusion rates in groups 2 through 4 in comparison to group 1. The complete occlusion rate was significantly lower in group 2 in comparison to group 1 (2.7 vs 20.1/1000 device-days, p = .003). There were no significant differences between the groups in the incidence of partial occlusion, time to the first day of catheter occlusion, phlebitis, or mechanical complication rate or severity. There was a trend toward a greater CRBSI rate in the positive-pressure saline group, though this finding did not reach statistical significance.


View this table:
[in this window]
[in a new window]

 
Table III Occlusion rates and other complications in each study group

 

The user satisfaction survey was administered to 132 nurses and was returned by 120 nurses, for an overall response rate of 90.9%. The results of these surveys are displayed in Figure 1. Scores for groups 3 and 4 were combined in analysis. Caregivers had experience with significantly more standard devices than either the valve or the positive-pressure-valve devices (p < .05). This finding is attributable to group 1 devices having been in use within the institution before the initiation of the study. There was significantly greater user satisfaction with the devices used in group 2, 3, and 4 in comparison to group 1's devices, as indicated by the responses to questions 3 through 7 (p < .05).


Figure 1
View larger version (25K):
[in this window]
[in a new window]

 
FIGURE 1. Results of the user satisfaction survey in all 4 study groups. Groups 3 and 4 were combined as the device was identical, with a different flush protocol. * p < .05.

 

    DISCUSSION
 Top
 METHODS
 RESULTS
 DISCUSSION
 CONCLUSIONS
 
Central venous catheters play a vital role in the care of children with both chronic and acute diseases. Though essential for high-quality care, these devices are not without risk. One of the more common complications associated with central venous catheters is occlusion. Catheter occlusion has been associated with delays in therapy, interruption in monitoring, and the need for treatments aimed at restoring patency.1,5,10 Unsuccessful attempts to restore catheter function often result in premature device removal.

There are many factors associated with central venous catheter occlusion. Peripherally placed catheter tips have significantly greater catheter occlusion complications than those that are centrally placed.11 Mechanical problems are common and generally associated with a catheter tip that resides next to an intracardiac structure, against the wall of a blood vessel, or in a peripheral location.1,12,13 Catheters may also become obstructed secondary to accumulation of blood products.6,14 Though manufacturers do not generally recommend the use of small catheters for blood sampling, they are often used for this purpose.15 A recent study noted that blood sampling through 3-Fr PICC devices is effective, though with an increased rate of occlusion.15

There are several approaches effective in the prevention of catheter occlusion. In addition to avoiding inappropriate tip placement and blood sampling, routine flushing with heparin or saline,1618 continuous infusion of maintenance fluids,19 continuous infusion of heparin-containing solutions,20 and clamping of the catheter or the extension tubing after catheter flushing21 have also been described. In addition, selection of an appropriate needleless connector device may be an important factor in preventing catheter occlusion. Despite claims supporting the importance of needleless connector devices, the use of these caps in prevention of catheter occlusion is one of the least-studied preventive strategies. Reflux of blood into the catheter tip may play a significant role in catheter occlusion. The role of various needleless connector device designs on prevention of reflux is unclear.6,21,22 There are a number of needleless connector devices available, including standard (no valve), single-valve, dual-valve, and positive-pressure-valve designs. The contribution of each of these connector designs in preventing blood reflux and occlusion has not been well studied. A recent investigation by our group compared 161 positive-pressure-valve devices to 151 standard devices in 153 children with central venous catheters.6 This study revealed significantly less complete occlusions with the positive-pressure-valve device than with the standard device (3.7% vs 11.9% respectively, p = .012). Positive-pressure-valve devices work by activating an antegrade flow of fluid through the catheter, thereby preventing reflux as the flush syringe is disconnected. There are few trials comparing single-valve or dual-valve devices to either standard or positive-pressure-valve devices.7,23 The primary purpose of our study was to broaden our understanding in needleless connector design and function by comparing the impact of standard, single-valve, and positive-pressure-valve needleless connector devices on central venous catheter occlusion rates.

Our results suggest a 3- to 9-fold lower complete occlusion rate in central venous catheters capped with either a single-valve or positive-pressure-valve device in comparison to a standard connector. These findings are very similar to a previous comparison between standard and positive-pressure-valve devices.6 We did not detect any significant difference between the complete occlusion rate in single- vs positive-pressure-valve devices. The complete occlusion protection afforded by valved connector devices is most likely secondary to the effect of these devices on minimizing accumulation of blood elements within the lumen of the catheter. Reflux of blood into the lumen after flushing was frequently noted when the standard device was used, despite a protocol to first flush, then clamp, and then disconnect the syringe. This same flush protocol was used with the single-valve needleless connector device. Clamping is likely the most important mechanism in eliminating blood reflux in both standard and single-valve devices. It is possible that flush-practice inconsistencies may have influenced the results within these 2 study arms. During the positive-pressure-valve device phase of the study, the flush protocol was changed to flush, disconnect, then clamp. This was necessary to allow the positive-pressure-valve device to activate, per manufacturer recommendations. This change in protocol was unlikely to have affected complete occlusion rates. Of note, the mean catheter size in group 1 subjects was smaller than group 3, and the mean catheter duration was greater in group 1 than all other groups. These 2 factors may have contributed to the differences in the primary outcome variable. However, when adjusted for catheter duration (events per 1000 device-days), the complete occlusion rate remained 2.1- to 7.4-fold lower in the single- and positive-pressure-valve device groups.

Despite the notable differences in complete occlusion rates, partial occlusion rates were not significantly different between any of the 3 device designs tested. Consistent with our previous findings,6 needleless cap design differences seem to afford little benefit in reducing the likely source of partial occlusions: blood-component accumulation on the external surface of the catheter.6,22

There is a paucity of data available comparing infection rates between different needleless connector designs. Casey et al24 recently demonstrated a reduced microbial contamination rate in positive-pressure-valve connectors compared to standard connectors. Though our data did not suggest any statistically significant differences between the groups, there was a notable 2-fold greater CRBSI rate in group 4 subjects vs those enrolled in the other groups. We estimate that 600 devices would need to be studied in each group for this difference to be adequately powered to detect a statistically significant difference with {alpha}= .05 and β= .7. The connector devices in group 4 were flushed with saline only, whereas devices in groups 1–3 were flushed with heparinized saline. The increased infection rate in the positive-pressure-device group flushed with saline in the current study was similar to the increased rate found in the positive-pressure-device group flushed with saline in our previous study.6 Heparin flush solutions contain preservatives with antimicrobial activity, which may have explained the differences in infection rates.25 In addition, prophylactic heparin has been noted to reduce the risk of catheter-related thrombosis, a factor associated with CRBSI.26 The use of heparin-containing flush solutions may play an important role in protection against catheter bacterial growth, and this subject will require further investigation.

Several factors are important in decisions to adopt new device technology into a health care organization, such as device safety, usability, reliability, cost, and user satisfaction. Clinician satisfaction was significantly higher with the use of both the single-valve and positive-pressure-valve devices compared to the standard device. Users rated the 2-valved connector devices higher in many areas, including the time required to attach the device, the simplicity of the device, syringe attachment and injection of fluids/medications, syringe attachment and removal of blood, and in attachment of IV infusion tubing. The results of the user satisfaction survey, when combined with complete occlusion rates and cost data, were instrumental in adopting the single-valve connector device at CCHMC. However, after this change was implemented across the institution, numerous user complaints were reported regarding the single-valve device. The most significant comments were related to retrograde blood return into the catheter and the external chamber of the connector which occurred after the procedural flush protocol. Because of these concerns, a follow-up user satisfaction survey was distributed to 243 individuals and returned by 195 (80.2% response rate). This new survey contained an additional statement: "I am satisfied with this product and find it easy to use and problem free." In response to this question, 46/195 (23.6%) strongly agreed, 77/195 (39.5%) agreed, 46/195 (23.6%) were neutral, 22/195 (11.3%) disagreed, and 4/195 (2%) strongly disagreed. The most commonly reported problems were (1) retrograde blood return into the catheter or needleless connector, (2) frequent need to change the connector device because of blood stasis in the device chamber, and (3) the need to flush catheters more frequently than standard protocol. These data indicated new concerns over the frequency of cap failure not detected in the original study survey and pointed out the need for continued monitoring of outcomes data subsequent to implementation of a new device.


    CONCLUSIONS
 Top
 METHODS
 RESULTS
 DISCUSSION
 CONCLUSIONS
 
In conclusion, single-valve and positive-pressure-valve needleless connector devices are effective in reducing complete occlusion rates in children with central venous catheters when compared with standard connector devices. The additional cost of these devices appears to be warranted in avoiding the difficulties associated with catheter malfunction, delays in care, and treatment costs.27 Blood reflux in central venous catheter lumens capped with the single-valve device used in this study remains to be an ongoing concern. The contribution of this problem to replacement of single-valve devices, infection rates, and cost requires further study before this device design can be recommended. Saline flush protocols may be associated with a greater infection rate than heparinized saline flush protocols, and further study is also required.

This study was supported by a research grant from Becton Dickinson.

Received for publication April 21, 2005. Accepted for publication December 5, 2005.

  1. Stephens LC, Haire WD, Kotulak GD. Are clinical signs accurate indicators of the cause of central venous catheter occlusion? JPEN J Parenter Enteral Nutr.1995; 19:75 –79.[Abstract/Free Full Text]
  2. Hoch JR. Management of the complications of long term venous access. Semin Vasc Surg.1997; 10:135 –143.[Medline] [Order article via Infotrieve]
  3. Jacobs BR. Central venous catheter occlusion and thrombosis. Crit Care Clin.2003; 19:489 –514.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  4. Rumsey KA, Richardson DK. Management of infection and occlusion associated with vascular access devices. Semin Oncol Nurs. 1995;11:174 –183.[CrossRef][Medline] [Order article via Infotrieve]
  5. Krzywda EA. Predisposing factors, prevention and management of central venous catheter occlusions. J Intraven Nurs.1999; 22(6 suppl):S11 –S17.[Medline] [Order article via Infotrieve]
  6. Jacobs BR, Schilling S, Doellman D, Hutchinson N, Rickey M, Nelson S. Central venous catheter occlusion: a prospective, controlled trial examining the impact of a positive-pressure valve device. JPEN J Parenter Enteral Nutr.2004; 28:113 –118.[Abstract]
  7. Rummel MA, Donnelly PJ, Fortenbaugh C. Clinical evaluation of a positive pressure device to prevent central venous catheter occlusion: results of a pilot study. Clin J Oncol Nurs.2001; 5:261 –265.[Medline] [Order article via Infotrieve]
  8. O'Grady NP, Alexander M, Dellinger EP, et al. Guidelines for the prevention of intravascular catheter-related infections: the Hospital Infection Control Practices Advisory Committee, Centers for Disease Control and Prevention. Pediatrics.2002; 110:e51 .[Abstract/Free Full Text]
  9. Garner IS, Jarvis WR, Emori TG, Horan TC, Hughes JM. CDC definitions for nosocomial infections. Am J Infect Control. 1988; 16:128 –140.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  10. Deitcher SR, Fesen MR, Kiproff PM, et al. Safety and efficiency of alteplase for restoring function in occluded central venous catheters: results of the cardiovascular thrombotic to open occluded lines trial. J Clin Oncol. 2002;20:317 –324.[Abstract/Free Full Text]
  11. Racadio JM, Doellman DA, Johnson ND, Bean JA, Jacobs BA. Pediatric peripherally inserted central catheters: complication rates related to catheter tip location. Pediatrics.2001; 107:E28 .[CrossRef][Medline] [Order article via Infotrieve]
  12. Fratino G, Molinari A, Mazzola C, et al. Prospective study of indwelling central venous catheter-related complications in children with Broviac or clampless valved catheters. J Pediatr Hematol Oncol. 2002;24:657 –661.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  13. Andris DA, Krzywda EA, Schulte W, Ausman R, Quebbeman EJ. Pinch off syndrome: a rare etiology for central venous catheter occlusions. JPEN J Parenter Enteral Nutr.1994; 18:531 –533.[Abstract/Free Full Text]
  14. Cassidy FP, Zajko AB, Bron KM, Reilly JJ, Peitzman AB, Steed DL. Noninfectious complications of long term central venous catheters: radiology evaluation and management. Am J Radiol.1987; 149:671 –675.[Abstract/Free Full Text]
  15. Knue M, Doellman D, Rabin K, Jacobs BR. The efficacy and safety of blood sampling through peripherally inserted central catheter devices in children. J Infus Nurs.2005; 28:30 –35.[Medline] [Order article via Infotrieve]
  16. Shah PS, Ng E, Sinha AK. Heparin for prolonging peripheral intravenous catheter use in neonates. Cochrane Database Syst Rev. 2004;4CD002774 .
  17. Rabe C, Gramann T, Sons X, et al. Keeping central venous lines open: a prospective comparison of heparin, vitamin C and sodium chloride sealing solutions in medical patients. Intensive Care Med. 2002;28:1172 –1176.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  18. Meyer BA, Little CJ, Thorp JA, Cohen GR, Yeast JD. Heparin versus normal saline as a peripheral line flush in maintenance of intermittent intravenous lines in obstetric patients. Obstet Gyncol. 1995;85:433 –436.[CrossRef]
  19. Terry J, Baranowski L, Lonsway RA, Hedrick C, eds. Intravenous Therapy: Clinical Principles and Practice. Philadelphia, PA: Saunders; 1995.
  20. Shah P, Shah V. Continuous heparin infusion to prevent thrombosis and catheter occlusion in neonates with peripherally placed percutaneous central venous catheters. Cochrane Database Syst Rev.2001; CD002772.
  21. Hadaway LC. Major thrombolytic and non-thrombolytic complications: loss of patency. J Intraven Nurs.1998; 21(5 suppl):S143 –S160.[Medline] [Order article via Infotrieve]
  22. Hoshal VL, Ause RG, Hoskins PA. Fibrin sleeve formation on indwelling subclavian central venous catheters. Arch Surg. 1991;102:353 –357.
  23. Cheeseman D. Intravenous care: the benefits of closed-system connectors. B J Nurs.2001; 10:287 –295.
  24. Casey AL, Worthington T, Lambert PA, Quinn D, Faroqui MH, Elliott TS. A randomized, prospective clinical trial to assess the potential infection risk associated with the Posiflow needleless connector. J Hosp Infect. 2003;54:288 –293.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  25. Elliott TS, Curran A. Effects of heparin and chlorbutol on bacterial colonization of intravascular cannulae in an in vitro model. J Hosp Infect.1989; 14:193 –200.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  26. Randolph AG, Cook DJ, Gonzales CA, Andrew M. Benefit of heparin in central venous and pulmonary artery catheters: a meta-analysis of randomized controlled studies. Chest.1998; 113:165 –171.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  27. Jacobs BR, Haygood M, Hingl J. Recombinant tissue plasminogen activator in the treatment of central venous catheter occlusion in children. J Pediatr.2001; 139:593 –596.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]

Journal of Parenteral and Enteral Nutrition, Vol. 30, No. 2, 85-90 (2006)
DOI: 10.1177/014860710603000285


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati    What's this?


This article has been cited by other articles:


Home page
JCOHome page
S. Cesaro, G. Tridello, M. Cavaliere, L. Magagna, P. Gavin, R. Cusinato, N. Zadra, G. Franco Zanon, L. Zanesco, and M. Carli
Prospective, Randomized Trial of Two Different Modalities of Flushing Central Venous Catheters in Pediatric Patients With Cancer
J. Clin. Oncol., April 20, 2009; 27(12): 2059 - 2065.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Saved Citations
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Request Reprints
Right arrow Add to My Marked Citations
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Right arrow Citing Articles via Scopus
Google Scholar
Right arrow Articles by Schilling, S.
Right arrow Articles by Jacobs, B. R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Schilling, S.
Right arrow Articles by Jacobs, B. R.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?