Sign In to gain access to subscriptions and/or personal tools.
|
Use of Triple-Lumen Subclavian Catheters for Administration of Total Parenteral Nutrition
Nena Clark-Christoff, RN, CNSN
Department of Surgery, Duke University Medical Center, Durham, North Carolina
Virginia A. Watters, RN, CNSN
Department of Surgery, Duke University Medical Center, Durham, North Carolina
Wanet Sparks, RN, CNSN
Department of Surgery, Duke University Medical Center, Durham, North Carolina
Phyllis Snyder, BA
Department of Surgery, Duke University Medical Center, Durham, North Carolina
John P. Grant, MD
Department of Surgery, Duke University Medical Center, Durham, North Carolina
This study evaluated the safety of triple vs single-lumen catheters in intravenous nutrition. Patients who were judged likely to benefit from a triple-lumen catheter were randomized to receive either a single-lumen catheter, with additional peripheral or central venous access as needed, or a triple-lumen catheter. All patients were at increased risk of catheter-related infection because of one or more of the following conditions: >60 years of age, breakdown of skin integrity, severe underlying illness, diagnosis of acute pancreatitis, recent head or neck surgery, or presence of a preexisting infection. Patients were excluded who had neutropenia, were immuno-suppressed, had body burns >40%, or had contaminated wounds in the subclavicular area. Of 204 patients entered between June 1989 and November 1991, 177 completed the required 7 days of therapy. Seventy-eight of these patients were randomized to a single-lumen catheter and 99 to a triple-lumen catheter. Catheters were inserted and maintained by the Nutrition Support Team. Dressings were monitored daily and changed weekly using a bio-occlusive dressing. When parameters were met for a possible septic episode, simultaneous peripheral and central catheter blood cultures were obtained using the Isolator method. Catheter-related sepsis was considered present if the colony count from a central catheter lumen was 5 times that of the peripheral blood. The incidence of catheter-related sepsis for single-lumen catheters was 2.6% (2 of 78) compared with 13.1% for triple-lumen catheters (13 of 99) (p < .01). No correlation was found with the number of insertion attempts, catheter days, or patient's age. We conclude that triple-lumen catheters should not be routinely used for intravenous nutrition because of an increased risk for central catheter-related sepsis. (Journal of Parenteral and Enteral Nutrition 16: 403-407, 1992)
Journal of Parenteral and Enteral Nutrition, Vol. 16, No. 5,
403-407 (1992)
DOI: 10.1177/0148607192016005403

CiteULike Complore Connotea Del.icio.us Digg Reddit Technorati Twitter What's this?
This article has been cited by other articles:

|
 |

|
 |
 
A.S.P.E.N. Board of Directors and Task Force on Pa, M. Kochevar, P. Guenter, B. Holcombe, A. Malone, and J. Mirtallo
A.S.P.E.N. Statement on Parenteral Nutrition Standardization
JPEN J Parenter Enteral Nutr,
September 1, 2007;
31(5):
441 - 448.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D. G. Maki, D. M. Kluger, and C. J. Crnich
The Risk of Bloodstream Infection in Adults With Different Intravascular Devices: A Systematic Review of 200 Published Prospective Studies
Mayo Clin. Proc.,
September 1, 2006;
81(9):
1159 - 1171.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. J. Sands
Vascular Access in the Adult Home Infusion Patient
JPEN J Parenter Enteral Nutr,
January 1, 2006;
30(1_suppl):
S57 - S64.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Zurcher, M. R. Tramer, and B. Walder
Colonization and Bloodstream Infection with Single- Versus Multi-Lumen Central Venous Catheters: A Quantitative Systematic Review
Anesth. Analg.,
July 1, 2004;
99(1):
177 - 182.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D. C. McGee and M. K. Gould
Preventing Complications of Central Venous Catheterization
N. Engl. J. Med.,
March 20, 2003;
348(12):
1123 - 1133.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
N. P. O'Grady, M. Alexander, E. P. Dellinger, J. L. Gerberding, S. O. Heard, D. G. Maki, H. Masur, R. D. McCormick, L. A. Mermel, M. L. Pearson, et al.
Guidelines for the Prevention of Intravascular Catheter-Related Infections
Pediatrics,
November 1, 2002;
110(5):
e51 - 51.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A man with a murmur requiring nutritional support
Postgrad. Med. J.,
August 1, 2001;
77(910):
537f - 537.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. E. Orr
Vascular Access Device Selection for Parenteral Nutrition
Nutr Clin Pract,
August 1, 1999;
14(4):
172 - 177.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
E. A. Krzywda, D. A. Andris, and C. E. Edmiston Jr
Catheter Infections: Diagnosis, Etiology, Treatment, and Prevention
Nutr Clin Pract,
August 1, 1999;
14(4):
178 - 190.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
C. Charalambous, S. M. Swoboda, J. Dick, T. Perl, and P. A. Lipsett
Risk Factors and Clinical Impact of Central Line Infections in the Surgical Intensive Care Unit
Arch Surg,
November 1, 1998;
133(11):
1241 - 1246.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D. N. Gilbert, R. J. Dworkin, S. R. Raber, and J. E. Leggett
Outpatient Parenteral Antimicrobial-Drug Therapy
N. Engl. J. Med.,
September 18, 1997;
337(12):
829 - 839.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. P. Grant
Invited Review: Nutrition Care of Patients With Acute and Chronic Respiratory Failure
Nutr Clin Pract,
February 1, 1994;
9(1):
11 - 17.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
J. J. Payne-James and H. T. Khawaja
Review: First Choice for Total Parenteral Nutrition: The Peripheral Route
JPEN J Parenter Enteral Nutr,
September 1, 1993;
17(5):
468 - 478.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Evans and M. A. Ryder
Invited Review: Vascular Access Devices: Perspectives on Designs, Complications, and Management
Nutr Clin Pract,
August 1, 1993;
8(4):
145 - 152.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
J. A. Kruse and N. J. Shah
Detection and Prevention of Central Venous Catheter-Related Infections
Nutr Clin Pract,
August 1, 1993;
8(4):
163 - 170.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
R. Wright, M. Fey, M.J. Zinn, and V. Ernst
Use of Triple-Lumen Subclavian Catheters for Administration of Total Parenteral Nutrition
JPEN J Parenter Enteral Nutr,
May 1, 1993;
17(3):
297 - 297.
[PDF]
|
 |
|
|
|